TECHNICAL SPECIFICATION |
Patient Response switch |
One push button |
Patient communication |
Talk Forward (TF) |
Input |
Tone, Warble Tone +5%, 5Hz (true sine wave frequency modulation). |
Output |
Left, Right, Bone L+R, Insert Phones, Insert Masking |
Dimensions (L×W×H) |
30x23x9cm / 12x9x4 inches. |
Weight |
1.3kg / 2.9lb |
Power Supply |
5VDC max 1.6A UE10 type only |
Batteries |
4×1.5V/1.2V Alkaline/NiMH Type AA Note: When the instrument is battery operated the maximum stimuli output level is reduced 20dB |
Operation environment Temperature |
Temperature: 15-35°C Relative Humidity: 30-90% Non-condensing Ambient pressure: 98-104 kPa |
Warm Up Time |
Approximately 1 Minute |
Internal storage |
500 Patient |
Data Connections (sockets) for connection of accessories |
1 x USB A for keyboard or printer 1 x USB B for PC connection (compatible with USB 1.1 and later) |
External devices (USB) |
Standard PC keyboard (for data entry) Supported printers: Please contact distributer for a list of approved PC printers |
Display |
4,3” (480×272) TFT color display |
Compatible software |
Diagnostic Suite – Noah OtoAccess and XML compatible |
Stimuli |
Tone |
125-8000Hz |
Warble Tone |
5Hz sine +/-% modulation |
Masking (Narrow band noise) |
IEC 60645-1 2012, 5/12 Octave filter with the same center frequency resolution as pure Tone Synchronous masking: Locks channel 2 attenuator to channel 1 attenuator |
Presentation |
Manual or Reverse. Single pulse. Multiple pulses 50-5000 msec. on/off |
Intensity |
AC: -10 to 120 dB HL BC: -10 to 80 dB Available Intensity Steps is 1, 2 or 5dB Extended range function: If not activated, the Air Conduction output will be limited to 20 dB below maximum output Extended range only available when mains powered |
Frequency range |
125Hz to 8kHz. (125Hz, 250Hz, 500 Hz, 750Hz, 1500Hz or 8kHz may freely be deselected) |